MicroPort Orthopedics Inc. recalls EVOLUTION MP Tibial Bases, Size 2+ LEFT, REF ETPKN2PL, knee tibial base

Recalled February 22, 2023 · FDA recall Z-0275-2023 · MicroPort Orthopedics

Hazard
One confirmed incident has been received that ETPKN2PL lot 1916559, size 2 Evolution MP Tibial Base, was opened during surgery and contained ETPKN7SL lot 1916715, size 7 Evolution MP Tibial Base, in the packaging. · FDA Class II
Units
24 units
Sold at
International Distribution to countries of: Germany, Italy, France, Belgium, Austria, Greece, Romania, Japan, and P.R. China.

Description

One confirmed incident has been received that ETPKN2PL lot 1916559, size 2 Evolution MP Tibial Base, was opened during surgery and contained ETPKN7SL lot 1916715, size 7 Evolution MP Tibial Base, in the packaging.

Product
EVOLUTION MP Tibial Bases, Size 2+ LEFT, REF ETPKN2PL, knee tibial base
Manufacturer(s)
MicroPort Orthopedics Inc.