Mindray DS USA, Inc. dba Mindray North America recalls Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00

Recalled October 24, 2018 · FDA recall Z-0170-2019 · Mindray DS USA, Inc.

Hazard
The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line at the centralized viewing station and will manifest if more than 64 TD60 transmitters have been connected to a given central station server. Each central station supports only 32 transmitters at a time. For the issue to manifest, the BeneVision DMS needs to be configured to use flexible device assignment, have more than 64 TD60 transmitters, and the workflow must allow 64 TD60s to move around between multiple central stations in the same care area. · FDA Class II
Units
5
Sold at
Distributed to accounts in CA, CO, FL, MA, NY, PA, PR, TX, and WV.

Description

The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line at the centralized viewing station and will manifest if more than 64 TD60 transmitters have been connected to a given central station server. Each central station supports only 32 transmitters at a time. For the issue to manifest, the BeneVision DMS needs to be configured to use flexible device assignment, have more than 64 TD60 transmitters, and the workflow must allow 64 TD60s to move around between multiple central stations in the same care area.

Product
Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00
Manufacturer(s)
Mindray DS USA, Inc. dba Mindray North America