Mindray DS USA, Inc. dba Mindray North America recalls Rosebud Vital Signs Monitor with NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5 - Produ
Recalled November 6, 2019 · FDA recall Z-0238-2020 · Mindray DS USA, Inc.
Hazard
NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function · FDA Class II
Units
9 units
Sold at
US Nationwide distributions.
Description
NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function
Product
Rosebud Vital Signs Monitor with NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5 - Product Usage: The monitor is intended for spot-check monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP), and Temperature (TEMP) on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians
Manufacturer(s)
Mindray DS USA, Inc. dba Mindray North America
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →