Home / Recalls / Medical Devices
Mindray DS USA, Inc. dba Mindray North America recalls TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00
Recalled February 14, 2018 · FDA recall Z-0554-2018 · Mindray DS USA, Inc.
Hazard
A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate. · FDA Class II
Units
203
Sold at
Distributed in 26 states: AL, CA, CO, FL, GA, IA, IN, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, PA, SD, TX, UT, WA, and WY; and Puerto Rico.
Description
A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate.
Product
TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00
Manufacturer(s)
Mindray DS USA, Inc. dba Mindray North America
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →