Mizuho OSI recalls Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.

Recalled February 20, 2013 · FDA recall Z-0810-2013 · Mizuho OSI

Hazard
Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top. · FDA Class II
Units
869 devices
Sold at
USA Distribution -- Worldwide Distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and WY., and the countries of Australia, Canada, Egypt, Finland, France, Germany, China, Israel, Italy, Japan, Netherlands, New Zealand, P.R. China, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Taiwan R.O.C., UAE and UK.

Description

Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.

Product
Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.
Manufacturer(s)
Mizuho OSI

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov