RecallWatchUSA

Home / Recalls / Medical Devices

Mobius Imaging, LLC recalls MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography (CT) system, software version 2.1.1

Recalled December 7, 2022 · FDA recall Z-0344-2023 · Mobius Imaging

Hazard
Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing. · FDA Class II
Units
3
Sold at
Domestic distribution to consignees in Colorado, Tennessee, and Utah.

Description

Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing.

Product
MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography (CT) system, software version 2.1.1
Manufacturer(s)
Mobius Imaging, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Mobius Imaging, LLC recalls MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography (CT) system, software version 2.1.1 | RecallWatchUSA