Home / Recalls / Medical Devices
Motion Concepts recalls TRx Vertical Positioner, Classification Name: Wheelchair Component - Power Positioning System.
Recalled September 24, 2014 · FDA recall Z-2657-2014 · Motion Concepts
Hazard
Motion Concepts identified that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the Vertical Position System (TRx Vertical Positioner). · FDA Class II
Units
US: 59 units
Sold at
US Distribution in states of: AR, CA, CT, DE, IA, ID, IL, IN, MI, MD, MO, NE, NY, OH, PA, TN, TX, and WA.
Description
Motion Concepts identified that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the Vertical Position System (TRx Vertical Positioner).
Product
TRx Vertical Positioner, Classification Name: Wheelchair Component - Power Positioning System.
Manufacturer(s)
Motion Concepts
Made in
Canada
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →