Home / Recalls / Medical Devices
Nakanishi Inc. recalls Diamond Coated Tips and Burs are accessories to the air powered dental handpiece (S-MAX pico) The intended use of the t
Recalled September 7, 2016 · FDA recall Z-2674-2016 · Nakanishi
Description
Through an error, the labeling failed to include the symbol or other text indicating that the diamond coated tips were single use only. A contributing factor to this error is that in other markets such as Japan, the diamond coated tips are multiple use and are not limited to single use only. The tips are stainless steel with a coating of fine diamond powder. The tips are sold non-sterile and are to be sterilized prior to first use. The tips were cleared as single use only. This problem was discovered during an ongoing, proactive, quality system improvement plan that was initiated by Nakanishi in October 2014. Because the correct instructions for sterilization are not available, it is possible that the devices might not be sterile prior to use. The absence of the single use statement introduces a risk of cross-contamination if the tips are reprocessed and reused.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →