Home / Recalls / Medical Devices
Natus Manufacturing Limited recalls Natus Warmette Blanket Cabinet and Warmette Dual Cabinet The Natus Warmette Blanket Cabinet is designed to store and wa
Recalled January 18, 2017 · FDA recall Z-0951-2017 · Natus Manufacturing
Hazard
The fuse blows during initial power up or after a period of time. This results in a complete loss of power to the Cabinet and in some cases has resulted in the fuse cover for the power inlet melting slightly. · FDA Class II
Units
31
Sold at
Worldwide Distribution - US including NJ, NC, TX, IL, VA, NY, FL, CA, MS, WA, IN, and MA. and Internationally to Canada and Costa Rica.
Description
The fuse blows during initial power up or after a period of time. This results in a complete loss of power to the Cabinet and in some cases has resulted in the fuse cover for the power inlet melting slightly.
Product
Natus Warmette Blanket Cabinet and Warmette Dual Cabinet The Natus Warmette Blanket Cabinet is designed to store and warm blankets and hospital linens; while, the Natus Warmette Dual Cabinet is designed to store and warm blankets and hospital linens in one compartment and fluids (irrigation and/or injection) in the other compartment within the storage times and warming temperatures recommended by the manufacturers of those items
Manufacturer(s)
Natus Manufacturing Limited
Made in
Ireland
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →