RecallWatchUSA

Home / Recalls / Medical Devices

Natus Medical Incorporated recalls Natus neoBLUE blanket LED Phototherapy system which includes the following items: 006224 neoBLUE blanket LED Photothe

Recalled April 22, 2015 · FDA recall Z-1412-2015 · Natus Medical Incorporated

Hazard
neoBLUE blanket LED Phototherapy System is recalled because Natus has received feedback from some customers about early failure of the neoBLUE blanket system fiberoptic pads. These failures involve discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box. · FDA Class II
Units
2611 units (1911 units in the US and 700 units outside the US)
Sold at
distributed worldwide

Description

neoBLUE blanket LED Phototherapy System is recalled because Natus has received feedback from some customers about early failure of the neoBLUE blanket system fiberoptic pads. These failures involve discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.

Product
Natus neoBLUE blanket LED Phototherapy system which includes the following items: 006224 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS
Manufacturer(s)
Natus Medical Incorporated

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Natus Medical Incorporated recalls Natus neoBLUE blanket LED Phototherapy system which includes the following items: 006224 neoBLUE blanket LED Photothe | RecallWatchUSA