Natus Medical Incorporated recalls Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. Neonatal phototh

Recalled December 31, 2014 · FDA recall Z-0847-2015 · Natus Medical Incorporated

Hazard
Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the original device. While some customers may prefer using these lights at the higher intensity levels, some are concerned that patients may receive more intensity than desired. · FDA Class II
Units
1907 total (1413 in US and 494 international)
Sold at
worldwide

Description

Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the original device. While some customers may prefer using these lights at the higher intensity levels, some are concerned that patients may receive more intensity than desired.

Product
Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. Neonatal phototherapy.
Manufacturer(s)
Natus Medical Incorporated

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov