Home / Recalls / Medical Devices
Natus Neurology DBA Excel Tech., Ltd. (XLTEK) recalls Natus Quantum System with NeuroWorks Software. Catalog /Part Numbers: 013926. The Natus Quantum Amplifier is intended
Recalled January 13, 2016 · FDA recall Z-0581-2016 · Natus Neurology
Hazard
During an internal testing, Natus Medical Incorporated, identified that in Neuro Works 8.1 with Quantum hardware, the incoming signal is displayed with reversed polarity. · FDA Class II
Units
56 (43 US, 13 OUS)
Sold at
Worldwide Distribution: US Distribution including states of: AL, AZ, AR, CO, FL, GA, MD, MA, MN, MS, NJ, NY, OH, SC, TN, TX and UT; and countries of: INDIA, KOREA, SPAIN, SWITZERLAND, THAILAND, UNITED KINGDOM, and CANADA. .
Description
During an internal testing, Natus Medical Incorporated, identified that in Neuro Works 8.1 with Quantum hardware, the incoming signal is displayed with reversed polarity.
Product
Natus Quantum System with NeuroWorks Software. Catalog /Part Numbers: 013926. The Natus Quantum Amplifier is intended to be used as an electroencephalograph: to acquire, display, store and archive electrophysiological signals. The amplifier should be used in conjunction with Natus NeuroWorks /SleepWorks software to acquire scalp and intracranial electroencephalographic (EEG) signals as well as polysomnographic (PSG) signals.
Manufacturer(s)
Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
Made in
Canada
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →