RecallWatchUSA

Home / Recalls / Medical Devices

Natus Neurology Inc recalls NicoletOne Software v5.94, Catalog/Part Number: 482-649600. The NicoletOne data acquisition and review system, either w

Recalled October 21, 2015 · FDA recall Z-0117-2016 · Natus Neurology

Hazard
When using the NicoletOne v5.94 software, after exiting the impedance check function and returning to the EEG screen, the impedance check signal remains active in waveform, obscuring the EEG signals. · FDA Class II
Units
101 (88 US, 13 OUS)
Sold at
Worldwide distribution. US nationwide, INDIA, ROMANIA, DENMARK, JAPAN, AUSTRALIA, RUSSIA, BANGLADESH, SRI LANKA, BRUNEI, and UNITED ARAB EMIRATES.

Description

When using the NicoletOne v5.94 software, after exiting the impedance check function and returning to the EEG screen, the impedance check signal remains active in waveform, obscuring the EEG signals.

Product
NicoletOne Software v5.94, Catalog/Part Number: 482-649600. The NicoletOne data acquisition and review system, either with or without synchronous digital video. The system is intended for medical purposes to record, measure, store and display full band (FbEEG) cerebral EEG and extracerebral activity for Clinical EEG, Electrocorticography (EcOG), Long Term Monitoring (LTM), Intensive Care Unit (ICU) monitoring and Polysomnography (PSG) Sleep studies. While the Nicolet Neurodiagnostic systems are capable of displaying signals, such as Sp02 and EKG, the system is NOT intended for monitoring such signals for the preservation of life, The Nicolet Neurodiagnostic systems are intended to acquire, analyze, and display data
Manufacturer(s)
Natus Neurology Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Natus Neurology Inc recalls NicoletOne Software v5.94, Catalog/Part Number: 482-649600. The NicoletOne data acquisition and review system, either w | RecallWatchUSA