Home / Recalls / Medical Devices
Navajo Manufacturing Company recalls Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd
Recalled April 1, 2026 · FDA recall Z-1588-2026 · Navajo Manufacturing
Hazard
If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property. · FDA Class I
Units
28,457
Sold at
US Nationwide distribution in the states of CA, MS, FL, IN, LA, VA, TX.
Description
If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.
Product
Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd
Manufacturer(s)
Navajo Manufacturing Company
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →