Navilyst Medical, Inc., an AngioDyamics Company recalls Vaxcel PICC with PASV MST-30 Kit under the following labels: 1) 4F, UPN M001454570, Rx ONLY, 2) 5F, UPN M001454620 & UPN

Recalled June 22, 2016 · FDA recall Z-1969-2016 · Navilyst Medical, Inc., an AngioDyamics

Hazard
PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter. · FDA Class II
Units
1,786 units
Sold at
Worldwide Distribution US (Nationwide) Internationally to AU, GR, SE, SA, IN, CA, GB, BR, BE, NL, ES, KR, HK, AE, IT and ES.

Description

PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.

Product
Vaxcel PICC with PASV MST-30 Kit under the following labels: 1) 4F, UPN M001454570, Rx ONLY, 2) 5F, UPN M001454620 & UPN M001454720, Rx ONLY & 3) 6F, UPN M001454770, Rx ONLY. The Vaxcel PICC with PASV is indicated for use in establishing peripheral access to the central venous system for administration of fluids including, but not limited to, hydration agents, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. It is also indicated for blood specimen withdrawal. This product is intended for central venous access in adults, children and infants who require intravenous (IV) therapy.
Manufacturer(s)
Navilyst Medical, Inc., an AngioDyamics Company