Home / Recalls / Medical Devices
Navilyst Medical, Inc recalls PICC Convenience Kit with Accessories, BioFlo PICC with ENDEXO and PASV Valve Technology, 4F-55 cm SINGLE LUMEN, UPN Pro
Recalled October 15, 2014 · FDA recall Z-0036-2015 · Navilyst Medical
Hazard
NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes. · FDA Class II
Units
3 units/pouches
Sold at
Nationwide Distribution
Description
NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.
Product
PICC Convenience Kit with Accessories, BioFlo PICC with ENDEXO and PASV Valve Technology, 4F-55 cm SINGLE LUMEN, UPN Product No. H96560M0224751, REF/Catalog No. 60M022475, STERILE --- Indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood and for power injection of contrast media.
Manufacturer(s)
Navilyst Medical, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →