RecallWatchUSA

Home / Recalls / Medical Devices

NeuroLogica Corporation recalls Samsung GM85 Mobile Digital X-ray Imaging System Product Usage: The GM85 Digital X-ray imaging System is intended f

Recalled June 5, 2019 · FDA recall Z-1691-2019 · NeuroLogica

Hazard
Potential issue related to the operation of the safety latch within the column that supports the arm due to a possibility of potential bodily harm due to a fall of the arm in an unlikely event that the column wire breaks. · FDA Class II
Units
288 systems
Sold at
US Nationwide Distribution

Description

Potential issue related to the operation of the safety latch within the column that supports the arm due to a possibility of potential bodily harm due to a fall of the arm in an unlikely event that the column wire breaks.

Product
Samsung GM85 Mobile Digital X-ray Imaging System Product Usage: The GM85 Digital X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications
Manufacturer(s)
NeuroLogica Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov