RecallWatchUSA

Home / Recalls / Medical Devices

Neurotherm, Inc. recalls Neurotherm Curved Radiofrequency (RF) cannnula 10 cm 18 gauge, 10 mm active tip Model Number: C-1010-R-18 Product

Recalled March 30, 2016 · FDA recall Z-1247-2016 · Neurotherm

Hazard
Straight needle labeled as a curved needle · FDA Class II
Units
135 cannulas (27 boxes/5)
Sold at
US Nationwide Distribution in the states of CA ,CO, OH, OR

Description

Straight needle labeled as a curved needle

Product
Neurotherm Curved Radiofrequency (RF) cannnula 10 cm 18 gauge, 10 mm active tip Model Number: C-1010-R-18 Product Usage: The device is a radiofrequency needle, used for the application of radiofrequency energy with as intended result, thermo coagulation of nervous tissue in the human body for pain relief. They are designed to be used to treat chronic pain. During nerve ablation, the needle is used in conjunction with a radiofrequency generator, an electrode and grounding pads.
Manufacturer(s)
Neurotherm, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov