Home / Recalls / Medical Devices
Newport Corp recalls Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as
Recalled June 12, 2024 · FDA recall Z-1946-2024 · Newport
Hazard
According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern. · FDA Class II
Units
19
Sold at
US Nationwide
Description
According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.
Product
Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone.
Manufacturer(s)
Newport Corp
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →