Home / Recalls / Medical Devices
Nidek Inc recalls YC-1800 OPHTHALMIC YAG LASER SYSTEM; The YC-1800 allows the safe and bloodless treatment of intraocular diseases for
Recalled April 8, 2015 · FDA recall Z-1367-2015 · Nidek
Hazard
Dear Doctor letter and Quick Reference Guide is being sent to users of the YC-1800 to reiterate important safety and cleaning information after complaints of pitting lens that may cause unclear images. · FDA Class II
Units
799 units
Sold at
Nationwide Distribution.
Description
Dear Doctor letter and Quick Reference Guide is being sent to users of the YC-1800 to reiterate important safety and cleaning information after complaints of pitting lens that may cause unclear images.
Product
YC-1800 OPHTHALMIC YAG LASER SYSTEM; The YC-1800 allows the safe and bloodless treatment of intraocular diseases for outpatients by mechanical disruption caused by energy of the YAG laser pulses.
Manufacturer(s)
Nidek Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →