RecallWatchUSA

Home / Recalls / Medical Devices

Novarad Corporation recalls NovaPACS Diagnostic Viewer. NovaPACS is a picture archiving and communication system software that retrieves, distribute

Recalled September 23, 2020 · FDA recall Z-2956-2020 · Novarad

Hazard
The firm received a report of an atypical dataset being generated. When using the cross-localization feature and images from a modality that generates asymmetrical images, the cross-localization reference line may be inaccurately placed on any of the corresponding images that are open. · FDA Class II
Units
197 units
Sold at
Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Canada, Colombia, Dominican Republic, Indonesia, Jordan, Mexico, Peru, Phillipines, and United Kingdom.

Description

The firm received a report of an atypical dataset being generated. When using the cross-localization feature and images from a modality that generates asymmetrical images, the cross-localization reference line may be inaccurately placed on any of the corresponding images that are open.

Product
NovaPACS Diagnostic Viewer. NovaPACS is a picture archiving and communication system software that retrieves, distributes, and display images and data from different modalities including CT, MR, US, CR/DX, NM, PT, and XA.
Manufacturer(s)
Novarad Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Novarad Corporation recalls NovaPACS Diagnostic Viewer. NovaPACS is a picture archiving and communication system software that retrieves, distribute | RecallWatchUSA