NxStage Medical Inc recalls NxStage PureFlow-B Solution, Premixed Dialysate for Hemodialysis, Device Catalog Numbers RFP-400, RFP-401, RFP-402, RFP-

Recalled June 26, 2024 · FDA recall Z-2160-2024 · NxStage Medical

Hazard
The smaller chamber of the two chamber bag is bursting when nurses are attempting to mix the two chambers prior to use. The electrolyte fluid in the small chamber of the NxStage PureFlow B Solution has a pH of less than 2.0. The solution can be harmful if it comes in contact with the eyes or skin, and/or as a result of a slip or fall injury due to spilled fluid. · FDA Class II
Units
1316936 units
Sold at
US Nationwide distribution.

Description

The smaller chamber of the two chamber bag is bursting when nurses are attempting to mix the two chambers prior to use. The electrolyte fluid in the small chamber of the NxStage PureFlow B Solution has a pH of less than 2.0. The solution can be harmful if it comes in contact with the eyes or skin, and/or as a result of a slip or fall injury due to spilled fluid.

Product
NxStage PureFlow-B Solution, Premixed Dialysate for Hemodialysis, Device Catalog Numbers RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-405, RFP-406, RFP-453, RFP-454, RFP-456
Manufacturer(s)
NxStage Medical Inc