RecallWatchUSA

Home / Recalls / Medical Devices

O-Two Medical Technologies, Inc. recalls e700 Automatic Transport Ventilator, Model Number 01EVE700

Recalled April 10, 2019 · FDA recall Z-0978-2019 · O-Two Medical Technologies

Hazard
There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit. · FDA Class I
Units
632
Sold at
The products were distributed to the following US states: AZ, CA, CO, FL, OH, PA, TX, and VA.

Description

There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.

Product
e700 Automatic Transport Ventilator, Model Number 01EVE700
Manufacturer(s)
O-Two Medical Technologies, Inc.
Made in
Canada

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov