RecallWatchUSA

Home / Recalls / Medical Devices

Ohmeda Medical recalls Giraffe Exam Light The Exam Light illuminates body surfaces and facilitates patient observation during a medical examin

Recalled March 29, 2017 · FDA recall Z-1501-2017 · Ohmeda Medical

Hazard
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury. · FDA Class II
Units
5,993 (5,080 US; 913 OUS)
Sold at
Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,

Description

Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.

Product
Giraffe Exam Light The Exam Light illuminates body surfaces and facilitates patient observation during a medical examination. It increases ambient light to facilitate performance of procedures on the mattress field. It is designed to mount on the accessory railings of the Giraffe OmniBed, incubators and warmers except for Panda and Giraffe Warmers. It is not intended for use to support surgical field lighting.
Manufacturer(s)
Ohmeda Medical

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Ohmeda Medical recalls Giraffe Exam Light The Exam Light illuminates body surfaces and facilitates patient observation during a medical examin | RecallWatchUSA