Home / Recalls / Medical Devices
Ohmeda Medical recalls Giraffe OmniBed Ohmeda Medical Omnibed- The OmniBed is a combination of an infant incubator ( incubator ) and an infant
Recalled March 29, 2017 · FDA recall Z-1509-2017 · Ohmeda Medical
Hazard
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury. · FDA Class II
Units
5,993 (5,080 US; 913 OUS)
Sold at
Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,
Description
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
Product
Giraffe OmniBed Ohmeda Medical Omnibed- The OmniBed is a combination of an infant incubator ( incubator ) and an infant radiant warmer ( warmer ). Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment
Manufacturer(s)
Ohmeda Medical
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →