Olympus America Inc. recalls Olympus Endoscopic Flushing Pump, Model OPF-2. Facilitates washing of tissues to remove blood, feces and other organi
Recalled November 27, 2013 · FDA recall Z-0329-2014 · Olympus America
Hazard
There is a possibility that the pump head was not shipped in the locked position. If the pump head is in an unlocked position this may, under certain circumstances, cause the pump to stop and go into power stand-by mode. · FDA Class II
Units
263
Sold at
Worldwide distribution: USA (nationwide) and the country of Canada.
Description
There is a possibility that the pump head was not shipped in the locked position. If the pump head is in an unlocked position this may, under certain circumstances, cause the pump to stop and go into power stand-by mode.
Product
Olympus Endoscopic Flushing Pump, Model OPF-2. Facilitates washing of tissues to remove blood, feces and other organic matter during endoscopic procedures.
Manufacturer(s)
Olympus America Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →