Olympus Corporation of the Americas recalls Colonoscope, Model Number CF-HQ190I.

Recalled March 6, 2024 · FDA recall Z-1206-2024 · Olympus Corporation of the Americas

Hazard
Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads. · FDA Class II
Units
1
Sold at
US Nationwide distribution.

Description

Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.

Product
Colonoscope, Model Number CF-HQ190I.
Manufacturer(s)
Olympus Corporation of the Americas

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov