Olympus Corporation of the Americas recalls EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Number Q180-AC

Recalled November 15, 2023 · FDA recall Z-0190-2024 · Olympus Corporation of the Americas

Hazard
There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope. · FDA Class I
Units
1154 devices
Sold at
US Nationwide distribution

Description

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

Product
EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Number Q180-AC
Manufacturer(s)
Olympus Corporation of the Americas