Home / Recalls / Medical Devices
Olympus Corporation of the Americas recalls EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190, BF-H190, BF-Q190 and BRONCHOVIDEOSCOPE OLYMPUS BF-Q170 Model
Recalled November 15, 2023 · FDA recall Z-0192-2024 · Olympus Corporation of the Americas
Hazard
There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope. · FDA Class I
Units
32211 devices
Sold at
US Nationwide distribution
Description
There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
Product
EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190, BF-H190, BF-Q190 and BRONCHOVIDEOSCOPE OLYMPUS BF-Q170 Model Number BF-Q170
Manufacturer(s)
Olympus Corporation of the Americas
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →