RecallWatchUSA

Home / Recalls / Medical Devices

Olympus Corporation of the Americas recalls Gyrus Acmi PKS Omni 5mm/33cm-intended for the electrosurgical coagulation, mechanical grasping, and dissection of tissue

Recalled September 8, 2021 · FDA recall Z-2380-2021 · Olympus Corporation of the Americas

Hazard
Expired adhesive used as an electrical insulation barrier and as a secondary retention bond failed and could result in foreign body / biologic reaction within the patient and / or electrical continuity to patient causing tissue burn to critical structures. · FDA Class II
Units
6 boxes (5 per/box)=30 units
Sold at
US Nationwide distribution in the states of CA, GA, MN, TN.

Description

Expired adhesive used as an electrical insulation barrier and as a secondary retention bond failed and could result in foreign body / biologic reaction within the patient and / or electrical continuity to patient causing tissue burn to critical structures.

Product
Gyrus Acmi PKS Omni 5mm/33cm-intended for the electrosurgical coagulation, mechanical grasping, and dissection of tissue during the performance of laparoscopic and general open surgical procedures when used in conjunction with the Gyrus ACMI G400 workstation Ref: 970010PC
Manufacturer(s)
Olympus Corporation of the Americas

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Olympus Corporation of the Americas recalls Gyrus Acmi PKS Omni 5mm/33cm-intended for the electrosurgical coagulation, mechanical grasping, and dissection of tissue | RecallWatchUSA