Home / Recalls / Medical Devices
Olympus Corporation of the Americas recalls Olympus POWERSEAL 5MM, 44CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0544CJDA. An electrosurgi
Recalled October 5, 2022 · FDA recall Z-1815-2022 · Olympus Corporation of the Americas
Hazard
An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment · FDA Class II
Units
44 units (5/box); 133 units (5/box) OUS
Sold at
Nationwide OUS: AUSTRALIA, CANADA , HONG KONG, CHINA, KOREA
Description
An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment
Product
Olympus POWERSEAL 5MM, 44CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0544CJDA. An electrosurgical device use in laparoscopic/minimally invasive or open surgical procedures.
Manufacturer(s)
Olympus Corporation of the Americas
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →