RecallWatchUSA

Home / Recalls / Medical Devices

Olympus Scientific Solutions Americas recalls Olympus Scientific Solutions Americas Analytical X-ray System. Delta, Delta Element, Delta Plus, DPO-4000-XX-U-EN-EN-P

Recalled August 12, 2015 · FDA recall Z-1991-2015 · Olympus Scientific Solutions Americas

Hazard
Higher than published scatter around the probe head when testing some samples. At the point the customer would interact directly with the instrument (Trigger position), most readings were lower. However, measurements around the probe head were higher particularly on samples know to be highly scattering (plastics, soils) with minimal increases on high density samples (steel, aluminum). · FDA Class II
Units
1047
Sold at
Nationwide Distribution.

Description

Higher than published scatter around the probe head when testing some samples. At the point the customer would interact directly with the instrument (Trigger position), most readings were lower. However, measurements around the probe head were higher particularly on samples know to be highly scattering (plastics, soils) with minimal increases on high density samples (steel, aluminum).

Product
Olympus Scientific Solutions Americas Analytical X-ray System. Delta, Delta Element, Delta Plus, DPO-4000-XX-U-EN-EN-PM-SP12, DPO-6000-XX-U-EN-EN-AP-SP10. Mobile X-ray System.
Manufacturer(s)
Olympus Scientific Solutions Americas

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Olympus Scientific Solutions Americas recalls Olympus Scientific Solutions Americas Analytical X-ray System. Delta, Delta Element, Delta Plus, DPO-4000-XX-U-EN-EN-P | RecallWatchUSA