Home / Recalls / Medical Devices
Olympus Scientific Solutions Americas recalls Olympus Scientific Solutions Americas Corporation (OSSA) Model: VANTA Analytical X-Ray System, X-Ray Fluorescence
Recalled August 9, 2017 · FDA recall Z-2760-2017 · Olympus Scientific Solutions Americas
Hazard
Instruments may, under certain specific circumstances listed below, fail to stop when a test is initiated and no sample is present (proximity shut down fails to end test). · FDA Class II
Units
146 units
Sold at
Worldwide Distribution - USA (nationwide) Distribution
Description
Instruments may, under certain specific circumstances listed below, fail to stop when a test is initiated and no sample is present (proximity shut down fails to end test).
Product
Olympus Scientific Solutions Americas Corporation (OSSA) Model: VANTA Analytical X-Ray System, X-Ray Fluorescence
Manufacturer(s)
Olympus Scientific Solutions Americas
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →