Home / Recalls / Medical Devices
Omega Medical Imaging, Inc. recalls Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085
Recalled November 1, 2023 · FDA recall Z-0156-2024 · Omega Medical Imaging
Hazard
An actuator separated from the pivot mechanism on a lower monitor boom system column and became detached and was restrained by the installed safety cable at one location. · FDA Class II
Units
52 units
Sold at
Worldwide distribution - US Nationwide and the country of Pakistan.
Description
An actuator separated from the pivot mechanism on a lower monitor boom system column and became detached and was restrained by the installed safety cable at one location.
Product
Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085
Manufacturer(s)
Omega Medical Imaging, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →