Home / Recalls / Medical Devices
Omega Medical Imaging, Inc. recalls Omega Medical Imaging Elevating Monitor Suspension Systems, Part Number 1000-0085
Recalled November 1, 2023 · FDA recall Z-0160-2024 · Omega Medical Imaging
Hazard
An actuator separated from the pivot mechanism on a lower monitor boom system column. The lower boom became detached at the point shown and was restrained by the installed safety cable. · FDA Class II
Units
31 units
Sold at
Alabama, California, Illinois, Indiana, Kentucky, Louisiana, Missouri, Montana, North Carolina, Ohio, Oklahoma
Description
An actuator separated from the pivot mechanism on a lower monitor boom system column. The lower boom became detached at the point shown and was restrained by the installed safety cable.
Product
Omega Medical Imaging Elevating Monitor Suspension Systems, Part Number 1000-0085
Manufacturer(s)
Omega Medical Imaging, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →