Home / Recalls / Medical Devices
OmniCell, Inc. recalls Omnicell IVX Station-For use in the automated preparation of pharmaceutical admixtures and syringes Model Number: 29892
Recalled May 31, 2023 · FDA recall Z-1617-2023 · OmniCell
Hazard
IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling · FDA Class II
Units
6 units
Sold at
AL, NY
Description
IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling
Product
Omnicell IVX Station-For use in the automated preparation of pharmaceutical admixtures and syringes Model Number: 298921373
Manufacturer(s)
OmniCell, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →