RecallWatchUSA

Home / Recalls / Medical Devices

Onkos Surgical, Inc. recalls ELEOS COLLAR STEM, CEMENTED, FLUTED, 15MM X 120MM. Model Number: HC-15120-03M. Component of Limb Salvage System with

Recalled June 4, 2025 · FDA recall Z-1818-2025 · Onkos Surgical

Hazard
Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems were mislabeled as size 13mm cemented stems. · FDA Class II
Units
23 units
Sold at
US Nationwide distribution in the states of California, Kentucky, and Wisconsin.

Description

Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems were mislabeled as size 13mm cemented stems.

Product
ELEOS COLLAR STEM, CEMENTED, FLUTED, 15MM X 120MM. Model Number: HC-15120-03M. Component of Limb Salvage System with BioGrip.
Manufacturer(s)
Onkos Surgical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Onkos Surgical, Inc. recalls ELEOS COLLAR STEM, CEMENTED, FLUTED, 15MM X 120MM. Model Number: HC-15120-03M. Component of Limb Salvage System with | RecallWatchUSA