RecallWatchUSA

Home / Recalls / Medical Devices

OraSure Technologies, Inc. recalls OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebo

Recalled August 24, 2016 · FDA recall Z-2403-2016 · OraSure Technologies

Hazard
Failed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C conditions. · FDA Class I
Units
3,075 units
Sold at
Distributed to one customer only in US state of Georgia.

Description

Failed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C conditions.

Product
OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola viruses.
Manufacturer(s)
OraSure Technologies, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

OraSure Technologies, Inc. recalls OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebo | RecallWatchUSA