RecallWatchUSA

Home / Recalls / Medical Devices

OriGen Biomedical, Inc. recalls The OriGen Reinforced Dual Lumen catheter (RDLC) is a duel lumen catheter with a wire reinforcement layer on the tip of

Recalled November 1, 2017 · FDA recall Z-0021-2018 · OriGen Biomedical

Hazard
Two users have reported to OriGen that a VV28F OriGen Reinforced Dual Lumen ECMO Catheter has had a separation of the clear extension tube from the hub connection. These adverse events resulted in patient injury (blood loss). · FDA Class I
Units
180 units
Sold at
US Nationwide and the countries of Austria, Belgium, Chile, Czech Republic, Italy, Kingdom of Saudi Arabia, Netherlands, Poland, Sweden, Thailand, Colombia, India, South Africa, United Arab Emirates.

Description

Two users have reported to OriGen that a VV28F OriGen Reinforced Dual Lumen ECMO Catheter has had a separation of the clear extension tube from the hub connection. These adverse events resulted in patient injury (blood loss).

Product
The OriGen Reinforced Dual Lumen catheter (RDLC) is a duel lumen catheter with a wire reinforcement layer on the tip of the device. A thin sheath is used to cover and secure the wire layer and is heat shrunk into a permanent position.
Manufacturer(s)
OriGen Biomedical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov