Ormco/Sybronendo recalls Spirit MB Bracket, UR Central +14T +5A.018, Part Number 494-0110, Lot Number 081351013. Product Usage: An orthodonti

Recalled April 9, 2014 · FDA recall Z-1330-2014 · Ormco/Sybronendo

Hazard
Ormco Corporation initiated this recall of Spirit MB Brackets, Part Number 494-0110, Lot Number 081351013, because the ID dot, which designates the bracket as distal/gingival, is misplaced. The misidentification may lead an orthodontist to place the bracket on the wrong tooth, leading to movement 10 degrees in the wrong direction. · FDA Class II
Units
150 units
Sold at
Internationally Distributed in the country of Japan.

Description

Ormco Corporation initiated this recall of Spirit MB Brackets, Part Number 494-0110, Lot Number 081351013, because the ID dot, which designates the bracket as distal/gingival, is misplaced. The misidentification may lead an orthodontist to place the bracket on the wrong tooth, leading to movement 10 degrees in the wrong direction.

Product
Spirit MB Bracket, UR Central +14T +5A.018, Part Number 494-0110, Lot Number 081351013. Product Usage: An orthodontic bracket. Spirit MB Brackets are a fixed attachment on a tooth which holds an archwire during orthodontic treatment.
Manufacturer(s)
Ormco/Sybronendo