Ortho-Clinical Diagnostics recalls Assay Data Disk (ADD), All DRV versions that support product within expiry dating Automates pre-analytical and post-a

Recalled June 17, 2015 · FDA recall Z-1729-2015 · Ortho-Clinical Diagnostics

Hazard
Software anomaly regarding urine samples that require acidified pretreatment. The software has allowed urine samples for assays that require acidified pretreatment to be metered from the same sample container as assays that required no pretreatment when the sample was programmed. · FDA Class II
Units
2,387 units Total (987 domestically & 1400 internationally)
Sold at
Worldwide Distribution - US Nationwide, Brazil, Chile, Mexico, Australia, Hong Kong, India, Taiwan, Belgium, Denmark, France, Ireland, Italy, Portugal, Romania, Spain, Sweden and the UK.

Description

Software anomaly regarding urine samples that require acidified pretreatment. The software has allowed urine samples for assays that require acidified pretreatment to be metered from the same sample container as assays that required no pretreatment when the sample was programmed.

Product
Assay Data Disk (ADD), All DRV versions that support product within expiry dating Automates pre-analytical and post-analytical sample and data management for in vitro diagnostic use. VITROS Urine Assays: VITROS Calcium (Ca), Magnesium (Mg), and Phosphorus (Phos) requires acidification pretreatment prior to urine assay testing
Manufacturer(s)
Ortho-Clinical Diagnostics

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov