Ortho-Clinical Diagnostics, Inc. recalls VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)

Recalled March 13, 2024 · FDA recall Z-1230-2024 · Ortho-Clinical Diagnostics

Hazard
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported. · FDA Class II
Units
15 US; 43 OUS
Sold at
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Description

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.

Product
VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
Manufacturer(s)
Ortho-Clinical Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov