RecallWatchUSA

Home / Recalls / Medical Devices

Ortho-Clinical Diagnostics, Inc. recalls VITROS XT 3400 Chemistry System, Catalog No. 6844458

Recalled March 13, 2024 · FDA recall Z-1227-2024 · Ortho-Clinical Diagnostics

Hazard
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported. · FDA Class II
Units
7 US; 4 OUS
Sold at
Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.

Description

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

Product
VITROS XT 3400 Chemistry System, Catalog No. 6844458
Manufacturer(s)
Ortho-Clinical Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Ortho-Clinical Diagnostics, Inc. recalls VITROS XT 3400 Chemistry System, Catalog No. 6844458 | RecallWatchUSA