Ortho-Clinical Diagnostics recalls VITROS 3600 Immunodiagnostic System, Catalog Number 6802783,for use in the in vitro quantitative, semi-quantitative, and

Recalled June 15, 2016 · FDA recall Z-1913-2016 · Ortho-Clinical Diagnostics

Hazard
Ortho Clinical Diagnostics confirmed a software timing anomaly in which the VITROS 3600 and 5600 Systems may unexpectedly aspirate from or dispense into an unintended sample that may lead to erroneous or contaminated results. · FDA Class II
Units
US: 218 units, Foreign: 587 units
Sold at
Nationwide, Puerto Rico and Foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.

Description

Ortho Clinical Diagnostics confirmed a software timing anomaly in which the VITROS 3600 and 5600 Systems may unexpectedly aspirate from or dispense into an unintended sample that may lead to erroneous or contaminated results.

Product
VITROS 3600 Immunodiagnostic System, Catalog Number 6802783,for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.
Manufacturer(s)
Ortho-Clinical Diagnostics

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov