Ortho-Clinical Diagnostics recalls VITROS 5,1 FS Chemistry System, Version 3.0 & below For use in the in vitro quantitative measurement of a variety of
Recalled December 26, 2018 · FDA recall Z-0651-2019 · Ortho-Clinical Diagnostics
Hazard
Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent. · FDA Class II
Units
1105
Sold at
US Nationwide Distribution
Description
Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.
Product
VITROS 5,1 FS Chemistry System, Version 3.0 & below For use in the in vitro quantitative measurement of a variety of analytes of clinical interest in certain biological fluids (depending on the analyte) such as serum, plasma, urine and cerebrospinal fluid.
Manufacturer(s)
Ortho-Clinical Diagnostics
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →