Ortho-Clinical Diagnostics recalls VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/300 SLDS, Product Code 1202670, UDI # 10758750008513 - Prod

Recalled June 24, 2020 · FDA recall Z-2364-2020 · Ortho-Clinical Diagnostics

Hazard
Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets. · FDA Class II
Units
73,071
Sold at
Worldwide distribution - US Nationwide including in the statesof Maine, Massachusetts, Vermont, Connecticut, New Jersey, New York, Pennsylvania, Washington DC, Maryland, Virginia, West Virginia, North Carolina, South Carolina, Georgia, Florida, Alabama, Tennessee, Missouri, Mississippi, Kentucky, Ohio, Indiana, Michigan, Iowa, Wisconsin, Minnesota, South Dakota, North Dakota, Montana, Illinois, Kansas, Nebraska, Louisiana, Arkansas, Oklahoma, Texas, Colorado, Idaho, Utah, Arizona, New Mexico, California, Hawaii, Oregon , Washington, and Alaska

Description

Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.

Product
VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/300 SLDS, Product Code 1202670, UDI # 10758750008513 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.
Manufacturer(s)
Ortho-Clinical Diagnostics

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov