Hazard
Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets. · FDA Class II
Units
64,697
Sold at
Worldwide distribution - US Nationwide including in the statesof Maine, Massachusetts, Vermont, Connecticut, New Jersey, New York, Pennsylvania, Washington DC, Maryland, Virginia, West Virginia, North Carolina, South Carolina, Georgia, Florida, Alabama, Tennessee, Missouri, Mississippi, Kentucky, Ohio, Indiana, Michigan, Iowa, Wisconsin, Minnesota, South Dakota, North Dakota, Montana, Illinois, Kansas, Nebraska, Louisiana, Arkansas, Oklahoma, Texas, Colorado, Idaho, Utah, Arizona, New Mexico, California, Hawaii, Oregon , Washington, and Alaska
Description
Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.
Product
VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/90 SLDS, Product Code 8112724, UDI # 10758750004393 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.
Manufacturer(s)
Ortho-Clinical Diagnostics
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →