Home / Recalls / Medical Devices
Ortho-Clinical Diagnostics recalls VITROS Chemistry Products Ca Slides, GEN 29, REF/Product Code 145 0261, Unique Device Identifier No. 10758750009114 or 2
Recalled August 3, 2016 · FDA recall Z-2273-2016 · Ortho-Clinical Diagnostics
Hazard
Ortho Clinical Diagnostics (OCD) confirmed a potential for biased calcium (Ca) results to be generated when using VITROS Ca Slides, GEN29, and the VITROS 250/350/950/5,1 FS, 4600 Chemistry System or the VITROS 5600 Integrated System. An increase in negatively biased calcium results complaints prompted the investigation that identified this product issue. · FDA Class II
Units
US: 27,625 sales units; Foreign: 4757 sales units
Sold at
Worldwide Distribution - US Nationwide in the countries of Puerto Rico and Foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, Poland and Venezuela.
Description
Ortho Clinical Diagnostics (OCD) confirmed a potential for biased calcium (Ca) results to be generated when using VITROS Ca Slides, GEN29, and the VITROS 250/350/950/5,1 FS, 4600 Chemistry System or the VITROS 5600 Integrated System. An increase in negatively biased calcium results complaints prompted the investigation that identified this product issue.
Product
VITROS Chemistry Products Ca Slides, GEN 29, REF/Product Code 145 0261, Unique Device Identifier No. 10758750009114 or 20758750009111, IVD -- Sales Unit: 300 slides/pack Product Usage: For in vitro diagnostic use only. VITROS Ca Slides quantitatively measure calcium (Ca) concentration in serum, plasma, and urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone disorders, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Manufacturer(s)
Ortho-Clinical Diagnostics
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →