RecallWatchUSA

Home / Recalls / Medical Devices

Ortho-Clinical Diagnostics recalls VITROS Chemistry Products URIC Slides, REF/Product Code 194 3927 (300 slides per sales unit), Unique Device Identifier

Recalled October 26, 2016 · FDA recall Z-0092-2017 · Ortho-Clinical Diagnostics

Hazard
Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays. · FDA Class II
Units
US: 18,749 sales units; Foreign: 94,928 sales units
Sold at
Worldwide Distribution - US (Nationwide) Puerto Rico and Bermuda), Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, and Venezuela.

Description

Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays.

Product
VITROS Chemistry Products URIC Slides, REF/Product Code 194 3927 (300 slides per sales unit), Unique Device Identifier No. 10758750000340, Rx ONLY, IVD --- For in vitro diagnostic use only.
Manufacturer(s)
Ortho-Clinical Diagnostics

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Ortho-Clinical Diagnostics recalls VITROS Chemistry Products URIC Slides, REF/Product Code 194 3927 (300 slides per sales unit), Unique Device Identifier | RecallWatchUSA