RecallWatchUSA

Home / Recalls / Medical Devices

Ortho-Clinical Diagnostics recalls VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITROS Systems: VITROS 5,1 FS Chemi

Recalled March 18, 2015 · FDA recall Z-1263-2015 · Ortho-Clinical Diagnostics

Hazard
In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanner unexpectedly skips sample cup positions within a sample tray. This issue only occurs when using the Batch Programming option; all other sample programming options function as intended. · FDA Class II
Units
Domestic: 62 units; Foreign: 10 units
Sold at
Worldwide Distribution-US (nationwide) and the countries of Canada, Australia and England.

Description

In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanner unexpectedly skips sample cup positions within a sample tray. This issue only occurs when using the Batch Programming option; all other sample programming options function as intended.

Product
VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITROS Systems: VITROS 5,1 FS Chemistry System, VITROS 3600 Immunodiagnostics System, VITROS 4600 Chemistry System, VITROS 5600 Integrated System. This product is an accessory for use with VITROS Analyzer Systems.
Manufacturer(s)
Ortho-Clinical Diagnostics

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Ortho-Clinical Diagnostics recalls VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITROS Systems: VITROS 5,1 FS Chemi | RecallWatchUSA